Medical Isolating Transformers
SIGA design and manufacture a wide range of transformers for the medical industry.
These can be supplied on either a stand alone unit or for assembly in customer’s equipment.
Medical Isolating transformers can be designed to meet client’s specific requirements, either toroidal or bobbin wound, and supplied either as the transformer only or assembled in a case.
Two standard models are also available, usually from stock – 750VA and 2000VA rating.
The specification for these is shown in the tables below:
Isolation Transformer Model EIT-750
Designed and built to comply with the requirements of Medical Equipment Specification, BS EN 60601-1
General Specification
- 750VA Medical Toroidal Isolation Transformer
- IEC inlet fused in each line
- Medical filter on input
- IEC Outlets 6 way
- Thermal Circuit Breaker illuminated
- Inrush current limiting
- Fused outputs
- Metal case 260mm W x 345mm D x 130mm H
- Carrying handles
Transformer Electrical Specification
- Primary : 0-230V 50/60Hz
- Secondary : 0-230V @750VA maximum
- Regulation : <4%
- Temperature Rise : <40°C
- Leakage : Primary to secondary <100µA @264V 50Hz
- Leakage : Primary to screen <500µA @ 264V 50Hz
- Flash Test: Primary to screen @2kV rms
- Flash Test : Primary to secondary @ 5kV rms
Overall Specification
- Input Voltage : 220V to 240V nominal 50/60Hz
- Input Power : 780W
- Input Filter : Medical low leakage <5µA rating 4A
- Input Fuses : Minature fuse 20mm x 5mm 6.3A anti surge type
- Output Rating : 220 to 240V @ 750VA maximum
- Output Fuses : Miniature fuse 20mm x 5mm 4A anti surge type
- Inrush Current : Thermistor limited 24A @264V
- Power On Indication : Incorporated in circuit breaker
- Output Indicator : Neon green inserted between fuses and output sockets
Download PDF Document here
SIGA (Electronics) Ltd
Isolation Transformer Model EIT-2000
Designed and built to comply with the requirements of Medical Equipment Specification, BS EN 60601-1
General Specification
- 2000VA Medical Toroidal Isolation Transformer
- IEC inlet fused in each line
- Medical filter on input
- IEC Outlets 6 way
- Thermal Circuit Breaker illuminated
- Inrush current limiting
- Fused outputs
- Metal case 260mm W x 345mm D x 130mm H
- Carrying handles
Transformer Electrical Specification
- Primary : 0-230V 50/60Hz
- Secondary : 0-230V @2000VA maximum
- Regulation : <4%
- Temperature Rise : <40°C
- Leakage : Primary to secondary <100µA @264V 50Hz
- Leakage : Primary to screen <500µA @ 264V 50Hz
- Flash Test: Primary to screen @2kV rms
- Flash Test : Primary to secondary @ 5kV rms
Overall Specification
- Input Voltage : 220V to 240V nominal 50/60Hz
- Input Power : 2100W
- Input Filter : Medical low leakage <5µA rating 10A
- Input Fuses : Miniature fuse 20mm x 5mm 10A anti surge type
- Output Rating : 220 to 240V @ 2000VA maximum
- Output Fuses : Miniature fuse 20mm x 5mm 10A anti surge type
- Inrush Current : Thermistor limited 24A @264V
- Power On Indication : Incorporated in circuit breaker
- Output Indicator : Neon green inserted between fuses and output sockets